2.6.7 Frequency Output Connections
The frequency output is used for a remote counter. It can be scaled to output a 1 to 10 kHz
signal proportional to mass or volume ow, temperature, pressure or density.
The frequency output requires a separate 5 to 36 VDC power supply and there are current
and power specications that must be observed when using this output. The output can
conduct a current up to 40 mA and can dissipate up to 200 mW. The output is isolated
from the meter electronics and power supply.
Freq. Out -
Freq. Out +
Freq. Out voltage = +V
Select resistor so that current
through Freq. Out <= 40 mA
Figure 21. Isolated Frequency Output Using External Power Supply
Freq. Out voltage = +V
Select resistor so that current through Freq. Out <= 40 mA
Freq. Out
+
Freq. Out -
Figure 22. Non-Isolated Frequency Output Using External Power Supply
2.6.5 Optional Backlight Connection
The loop power meter has an optional backlight connection provided. It is intended to be
powered by a separate 12 to 36 VDC at 35 mA max. power supply or by the pulse power
input. Both options are shown below.
Figure 23. Backlight Using External Power Supply