3.3.8 Fluid Menu
Password
ENTER
Use
keys to access menus.
Fluid
Menu
Flowing Fluid
Liquids >
Goyal-Dorais >
API 2540 >
Nat Gas AGA8 >
Real Gas >
Other Gas >
ENTER
Run Mode
< Liquid
Water
Ammonia
Chlorine
STD Press (PSIA)
xxxx
NORM Temp (C)
xxxx
NORM Press (KPA)
xxxx
STD Temp (F)
xxxx
Other Liquids >
< Mole Weight >
xxxx
< CRIT PRESS >
xxxx
< CRIT TEMP >
xxxx
< CRIT Z >
xxxx
< AL >
xxxx
< BL
xxxx
< Density @ 60F >
xxxx
< K0 >
xxxx
< K1 >
xxxx
< AL >
xxxx
< BL
xxxx
< Real Gas
Steam T & P Comp
< Specific Gravity >
xxxx
< MoleFract N2 >
xxxx
<MoleFract CO2>
xxxx
<Ref Temp(F)>
xxxx
< Ref Press(PSIA)
xxxx
< Specific Gravity >
xxxx
< Compress (Z) >
xxxx
< Viscosity
xxxx
Air
Argon
Ammonia
CO
CO2
Helium
Hydrogen
Methane
Nitrogen
Oxygen
< Density >
xxxx
< AL >
xxxx
< BL
xxxx
Select “Steam T & P Comp” for VT and
in the run mode screens.
“Sat Steam T Comp” for the fluid type
VTP models. The VT model will display
For a V model in any fluid, enter nominal operating temperature and
pressure as simulated values in the diagnostics menu.
Liquified Gas >
< Liquified Gas
Carbon Dioxide
Nitrogen
Hydrogen
Oxygen
Nitrous Oxide
Argon
Propane
Use the Fluid Menu to congure the ow meter for use with common gases, liquids and
steam. Your ow meter is pre-programmed at the factory for your application’s process
uid.
Reference Richard W. Miller, Flow Measurement Engineering Handbook (Third Edition,
1996), page 2-75 for denition and use of the Goyal-Doraiswamy equation and page 2-76 for
the denition and use of the API 2540 equation. Also, see Appendix B for Fluid Calculation
equations.
The units of measurement used in the Fluid Menu are preset and are as follows:
Mole Weight = lbm/(lbm•mol), CRIT PRESS = psia, CRIT TEMP = °R, Density = kg /m3 and
Viscosity = cP (centipoise).