F-2600 & F-2700 Vortex Flow Meter Manual 02/15 - 0808-7 / 19204 Page 99
SECTION 5.0: TROUBLESHOOTING AND REPAIR
5.1 HIDDEN DIAGNOSTICS MENUS
The menus shown on the following page can be accessed using the password 16363, then moving to
the display that reads “Diagnostics Menu” and pressing ENTER (rather than one of the arrow keys).
Use the right arrow key to move to the second level. Press EXIT to move from the second level back
to the rst. Press EXIT while in the rst level to return to the set-up menus.
Each of the menus on the following page will rst be dened followed by specic troubleshooting
steps.
!
WARNING
Before attempting any ow meter repair, verify that the line is not pressurized. Always remove
main power before disassembling any part of the mass ow meter.
!
CAUTION
Password 16363 will allow full access to the conguration and should be used carefully to avoid
changes that can adversely alter the function of the meter.